The FV1200MPE Exclusive is an optimized multiphoton only system. With a simplified scanhead with gold galvanometer mirrors for the highest multiphoton excitation efficiency, and high sensitivity GaAsP detectors, the FV1200MPE Exclusive provides a dedicated instrument for deep imaging in fixed and living tissue. Now supporting the new Insight laser from Spectra Physics and options such as the award winning SCALEVIEW objectives, reach the greatest imaging depths up to 8mm in SCALEVIEW-A2 cleared fixed tissue. High sensitivity detection and new high speed measurement software makes the system highly robust for applications from electrophysiology to large volume connectomics analysis.
SPECIFICATIONS
| B System | S System | T System | M System | ||
| Laser unit | IR pulsed laser with negative chirp for multiphoton excitation |
•Mode-locked Ti : sapphire laser [femtosecond laser (equipped with a group velocity compensation)], laser power unit, water-cooled circulating chiller |
|||
Visible light laser AOTF laser combiner |
LD laser: 405nm: 50mW, 440nm: 25mW, 473nm 15mW, 559nm 15mW, 635nm 20mW AOTF laser combiner Multi Ar laser (457nm, 488nm, 515nm, Total 30mW), HeNe (G) laser (543nm, 1mW) Modulation: Continuously adjustable via an AOTF (0.1 — 100% in 0.1% increments)
|
N/A | |||
| Single laser for visible light | LD473 laser (15mW) Depending on the type of modulation: light intensity modulation, shutter control, connected to the scanner via single-mode optical fiber. | N/A | |||
| Scanning Unit | Scanning method | Light deflection via 2 falvanometer scanning mirrors | |||
| Scanning modes | •Pixel size: 64 x 64 — 4096 x 4096 pixels Scanning speed: (pixel time): 2μs — 200μs High-speed scanning mode: 16 frames/s (256 x 256)
•Dimensions: Time, Z, (wavelength) (or any combination thereof) |
||||
| Zoom size | Observation position zoom with inclination width modification of galvanometer mirror: 1-50x (adjustable in 0.1x increments. | ||||
| Confocal detector (The M scanner does not have a confocal detector) |
|
N/A | |||
| Optics with infrared laser for multiphoton imaging |
|
||||
| Component incorporating the laser for multiphoton imaging | Main scanner for observation |
ASU scanner for laser light stimulation, Main Scanner for observation: VIS laser |
Incorporating 2 independent lasers for laser light stimulation/observation |
M scanner for observation |
|
| Detector for multiphoton imaging | Reflected light fluorescent detector | Photomultiplier (2 or 4 channels), Fluorescence wavelength can be selected with teh dedicated filter cube (replaceable) | |||
| Transmitted light fluorescent detector | Photomultiplier (2 channels),
fluorescent detector Fluorescence wavelength can be selected with the dedicated filter cube (replaceable) Exclusive equipment for the BX61WI upright microscope |
||||
| Transmitted DIC unit |
|
||||
| Z-drive |
|
||||
| Microscope | Upright microscopes: BX61WI, BX61 Inverted microscope: IX83 | ||||
| System control |
|
||||
| Required installation environment | Room temperature: 20 – 25°C, humidity: 75% or less@25°C, requires continuous (24-hour) power supply | ||||
| Vibration isolation table for microscope and laser installation, size | 1500mm x 1250mm |
1500 mm x1500 mm |
1700 mm x1700 mm |
1500mm x 1250mm |
|
* All other third party marks are property of their respective owners.